Stay tuned for our monthly news!

Stay informed

No, thanks

research

Article24 Nov 20267 min read

510(k) vs CE Marking: Why FDA clearance does not carry over to Europe (anymore)

A 510(k) does not carry over to Europe. Compare FDA clearance vs CE marking on time and cost, and learn how to fund the MDR runway. (131 characters)

510(k) vs CE marking: the short answer

What does a 510(k) give you in Europe?

A 510(k) gives you no legal right to sell in Europe. It does give you reusable evidence, and that distinction drives your budget.

The FDA question is narrow: is your device substantially equivalent to a legally marketed predicate? CDRH answers it on a stable clock. Decisions took about 141 days on average for FY2025, 144 for FY2024 and 141 for FY2023, per Medical Device User Fee Amendments (MDUFA) data through June 2026, counting FDA and submitter days.

The MDR asks a harder question: does your device meet the general safety and performance requirements on clinical evidence of its own? Nobody in Europe "clears" your device. You declare conformity, and for all but the lowest-risk devices a notified body designated by an EU country certifies your quality management system (QMS) and technical documentation.

Your US class does not predict your EU class: under MDR rules, a US class II device can land anywhere from class I to class III.

Evidence travels better: bench, biocompatibility, software, usability and risk files often survive a gap check. The clearance letter does not.

Why does treating CE marking as a second 510(k) fail?

It budgets a multi-quarter program like a short, fixed-fee event, and it starts the clock too late. The real FDA clearance vs CE mark gap sits in the cash-flow plan, not in the rules.

The clock starts with procurement. In Europe you first select, contract and pay your reviewer, a notified body. In a February 2026 survey for the European Commission, 51% of notified bodies put a new MDR certificate for QMS and product at 13 to 18 months after the written agreement. Another 31% said 19 to 24 months, and only 8% quoted 6 to 12.

Thirteen months, the low end of the common quote, is almost three times FDA's 141-day average, partly because the notified body also audits your QMS. Your runway does not care why. If your board expects EU revenue within 12 months and nothing is signed yet, at best 8% of notified bodies quote a timeline that fits. Even 24 months leaves little slack: 31% quote 19 to 24 months after signature.

The money behaves differently. FDA's FY2027 510(k) user fee is $28,653 for applications received from October 1, 2026, or $7,163 for small businesses. In Europe, manufacturers surveyed with data as of October 2025 report a median total cost of €100,000 for an initial MDR product certificate, against an average of €194,785. For an initial QMS certificate, the median was €50,000.

A fee and a reported total are not comparable, which is the point. With the average almost double the median, budget the median and reserve up to the average.

The predicate does not travel. On MDR equivalence vs predicate, CW1's reading of Article 61 is plain: a predicate rarely transfers. Substantial equivalence compares intended use and technology, while MDR equivalence must hold on technical, biological and clinical grounds, on data you can access. For implantable and class III devices, Article 61 generally requires clinical investigations. Equivalence to a competitor's device then needs a contract giving full, ongoing access to its technical documentation, and competitors have little reason to sign one.

MDSAP does not open the EU. FDA may accept Medical Device Single Audit Program (MDSAP) audit reports in place of routine inspections. The EU is only an observer and does not recognize MDSAP for CE marking. Yet 13 of the 15 MDSAP auditing organizations are EU notified bodies, according to a February 2025 paper by MedTech Europe and COCIR. Until EU law changes, there is no MDSAP CE marking shortcut.

How should a US team fund and sequence CE marking?

Run CE marking as a funded program with its own critical path, starting while the 510(k) is in review or just cleared.

  1. Audit the 510(k) file against the MDR. Confirm the EU class, then sort every document into reuse or build. The build pile, usually clinical evidence, sets budget and clock.
  2. Choose the clinical route before the notified body. Decide whether your own clinical data can carry the evaluation or a new investigation is needed. Plan as if competitor equivalence is unavailable.
  3. Run notified body procurement in parallel. Shortlist bodies designated for your device type and make the signed agreement a critical-path milestone. If your MDSAP auditing organization is also a notified body, ask to align audit calendars.
  4. Build a quarterly runway model. Give Europe its own lines: notified body fees, clinical evidence, an EU authorized representative and a person responsible for regulatory compliance. Add post-market clinical follow-up (PMCF) and periodic safety update reports (PSUR), then fund the whole curve before you sign.
  5. Sequence the launch with the certificate. Line up first markets and distributors during the review, so the certificate turns into revenue, not a second project.

Take a hypothetical company cleared in March 2027 that signs with a notified body in June 2027. On the common quote, certification lands between July 2028 and June 2029: five to nine quarters of EU spend after clearance, before any EU sale. The full route is in How to Get a CE Mark for a Medical Device After FDA Clearance in 2027.

510(k) vs CE marking: what does each difference cost a US team?

Each difference costs months of runway, a budget line, or both.

Dimension

FDA 510(k)

CE marking under MDR

What it means for a US team

Who reviews

FDA's CDRH, directly

A notified body you select, contract and pay

Procurement sits on the critical path

Legal standard

Substantial equivalence to a legally marketed predicate

Conformity with the MDR's general safety and performance requirements

You prove your device, not its likeness to another

Clinical evidence

Often bench and performance data, clinical data when needed

Clinical evaluation under Article 61. Class III and implantables generally need investigations

Budget it first: it moves cost and time most

Average time

About 141 days, FY2025 cohort, data through June 2026

13 to 24 months after the written agreement, per 82% of notified bodies, February 2026

About four to eight quarters, contract to certificate

Fee or reported cost

FDA user fee: $28,653 for FY2027, from October 2026, plus $13,785 annual registration

Reported total certificate cost: median €100,000 per initial product certificate, data as of October 2025

A fixed fee versus a program to fund

MDSAP recognition

Audit reports may replace routine inspections

Not recognized for CE marking. The EU is an observer

Expect a separate MDR audit

Post-market duties

Complaint handling, adverse event reporting, annual registration and listing

Post-market surveillance, PMCF, PSUR, vigilance, notified body surveillance audits

Recurring costs: operating budget, not project budget

More on post-clearance duties in EU MDR vs FDA: How Compliance Changes After Your Device Is Cleared.

How CW1 helps

CW1 audits your 510(k) file and EU plan, then designs the regulatory and financial strategy as one sequence. Our strategy and finance consultants work with doctors, so clinical and cash questions are answered together. We prepare the path and do the first regulatory and financial work with your team. Then your team continues alone, on a plan it owns.

Frequently asked questions

Does a 510(k) work in Europe?

No. The EU does not recognize FDA clearance, so you need CE marking under the MDR, though much of your test evidence can be reused.

What is the difference between FDA 510(k) clearance and CE marking under the MDR?

A 510(k) proves substantial equivalence to a predicate, with decisions taking about 141 days on average for the FY2025 cohort, per FDA data through June 2026. CE marking proves MDR conformity on your own clinical evidence, and 82% of notified bodies quoted 13 to 24 months in February 2026.

Can I use my 510(k) predicate as clinical equivalence for CE marking?

Rarely, in CW1's reading of MDR Article 61. For class III and implantable devices, equivalence to another manufacturer's device needs a contract giving full, ongoing access to its technical documentation.

Does an MDSAP certificate count toward CE marking?

No. The EU is only an observer to MDSAP and does not recognize it for CE marking, so plan a separate MDR audit.

How much does CE marking cost compared with a 510(k)?

FDA's FY2027 510(k) user fee, in force since October 2026, is $28,653, while manufacturers report a median total cost of €100,000 for an initial MDR product certificate, with data as of October 2025.

More insights

We use cookies to give you the best possible experience, and while some are essential, others help us understand how you use the site, so we can improve it. Click “Accept all cookies” to proceed as specified, “Decline optional cookies” to accept only essential cookies, or click “Manage my preferences” to choose what cookie types you will accept. Cookie Policy.

© CW1 Inc, All rights reserved